  • WSSX-100B 防爆电接点双金属温度计.jpg
  • WSSX-100B 隔爆型双金属温度计.jpg
  • WSSX-411B 防爆型电接点双金属温度计.jpg
  • WSSXB 隔爆型双金属温度计.jpg

WSSX-100B flameproof electric contact bimetal thermometer

Classification:

Bimetal thermometer

Key words:

Product Summary:

WSS-411B flameproof electric contact bimetal thermometers are designed and manufactured according to the national standard "electrical equipment for explosive gas environment" part 1: general requirements GB3836.1-2000, part 2: relevant regulations of flameproof "d" GB3836.2-2000 and the requirements of industry standard: "bimetal thermometer" JB/T8803-1998.

Tel: 15123315409

Return to List

Inquiry

Product Description

WSS-411B flameproof electric contact bimetal thermometers are designed and manufactured according to the national standard "electrical equipment for explosive gas environment" part 1: general requirements GB3836.1-2000, part 2: relevant regulations of flameproof "d" GB3836.2-2000 and the requirements of industry standard: "bimetal thermometer" JB/T8803-1998.
1. uses
Explosion-proof mark of flameproof electric contact bimetal thermometer: EXdⅡ CT6 (mark Ex, flameproof d, category T6). It is suitable for temperature measurement of medium with temperature lower than -50 500 ℃ in explosive gas environment. And through the use of electrical signal devices and electrical devices with corresponding explosion-proof performance, the measured temperature is automatically controlled, alarm, signal remote transmission and other functions, case protection level: IP65.
2. Instrument Structure and Explosion-proof Principle
The instrument is composed of an outer shell of flameproof structure and a bimetallic thermometer with electric contacts. During operation, when the combustible mixture inside the case explodes due to spark or arc (caused by the electrical signal device of the bimetallic thermometer with electric contacts), it will not cause an explosion outside the instrument, so as to achieve the purpose of explosion-proof.
3. basic parameters
1. See the following table for measuring range, accuracy level and electrical parameters of instrument





Shell Nominal
Diameter mm
Measurement range MPa Accuracy Level Electrical parameters
100 -40 80 ℃
-40 100 ℃;0~50 ℃
0~100 ℃;0~150 ℃
0~200 ℃;0~250 ℃
0~300 ℃;0~400 ℃



1.0;1.5

Rated power VA
Direct acting type 10
Magnetically assisted direct acting type 30

Maximum operating voltage V380

Maximum allowable current A
direct acting 0.7
Magnetically assisted direct acting 1.0

2. See Fig. 1 for overall dimensions

1